Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 111
Subsystem Commands
SOURce Subsystem
mode=0 or OFF disables the trigger output at the phase marker.
mode=1 or ON enables the trigger output at the phase marker.
Query response mode <newline>
mode is 0 or 1, and indicates that the trigger output at the phase marker is disabled or
enabled, respectively. Response data type is NR1.
Example :ARB:CURR:SIN:PMAR:STAT 0
:SOUR2:ARB:CURR:SIN:PMAR:STAT?
[:SOURce]:ARB:<CURRent|VOLTage>:SQUare:END:
TIME
Sets the end time of the current or voltage square waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:END:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:END:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter time Square waveform end time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SQU:END:TIME 0.1
:SOUR2:ARB:CURR:SQU:END:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:SQUare:STAR
t[:LEVel]
Sets the start level of the current or voltage square waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:STARt[:LEVel] data