4- 96 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SENSe Subsystem
Example :SENS:FUNC:STAT? “CURR”
:SENSe:REMote
Enables or disables the remote sensing. Remote sensing must be enabled to use the
4-wire connection (Kelvin connection).
Syntax :SENSe[c]:REMote mode
:SENSe[c]:REMote?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables remote sensing.
mode=1 or ON enables remote sensing.
Query response mode <newline>
mode is 0 or 1, and indicates that the remote sensing is off or on, respectively.
Response data type is NR1.
Example :SENS:REM 1
:SENS2:REM?
:SENSe:RESistance:OCOMpensated
Enables or disables the offset-compensated resistance measurement.
Syntax :SENSe[c]:RESistance:OCOMpensated mode
:SENSe[c]:RESistance:OCOMpensated?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables offset-compensated resistance measurement.
mode=1 or ON enables offset-compensated resistance measurement.
Query response mode <newline>
mode is 0 or 1, and indicates that offset-compensated resistance measurement is off
or on, respectively. Response data type is NR1.
Example :SENS:RES:OCOM 1