4- 124 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
response returns the present setting, INT1, INT2, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :ARB:CURR:UDEF:BOST:SIGN EXT3
:SOUR2:ARB:CURR:UDEF:BOST:SIGN?
[:SOURce]:ARB:<CURRent|VOLTage>:UDEFined:BO
STep:STATe
Enables or disables the trigger output for the user defined waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:UDEFined:BOSTep:STATe mode
[:SOURce[c]]:ARB:<CURRent|VOLTage>:UDEFined:BOSTep:STATe?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the trigger output.
mode=1 or ON enables the trigger output.
Query response mode <newline>
mode is 0 or 1, and indicates that the trigger output is disabled or enabled,
respectively. Response data type is NR1.
Example :ARB:CURR:UDEF:BOST:STAT 0
:SOUR2:ARB:CURR:UDEF:BOST:STAT?
[:SOURce]:ARB:<CURRent|VOLTage>:UDEFined[:LE
Vel]
Sets the current or voltage output data for each point of the user defined waveform
for the specified channel.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:UDEFined[:LEVel] list
[:SOURce[c]]:ARB:<CURRent|VOLTage>:UDEFined[:LEVel]?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.