4- 104 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
:SOUR2:ARB:CURR:EXP:TCON?
[:SOURce]:ARB:<CURRent|VOLTage>:EXPonent:TI
ME
Sets the time of the current or voltage exponential waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter time Time of the exponential waveform output, in seconds. value (0
to 1000)|MINimum|MAXimum|DEFault (default is 0).
Parameter data type is NRf+. Query does not support
time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:EXP:TIME 0.1
:SOUR2:ARB:CURR:EXP:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:RAMP:END[:
LEVel]
Sets the end level of the current or voltage ramp waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:END[:LEVel] data
[:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:END[:LEVel]? [data]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.