Keysight B2961A/B2962A SCPI Command Reference, Edition 6 3- 5
Common Commands
*ESE
*ESE
This command sets the bits of the Standard Event Status Enable register. This
command programs the Standard Event Status Enable register bits. The
programming determines which events of the Standard Event Status Enable register
are allowed to set the ESB (Event Summary Bit) of the Status Byte register. A 1 in
the bit position enables the corresponding event. For the SCPI status system, see
“Status System Diagram” on page 1-17.
Syntax *ESE enable_number
*ESE?
Parameter enable_number Decimal value that is the sum of the binary-weighted values for
the desired bits, hexadecimal, octal, or binary value. Parameter
data type is NR1 or NDN.
Query response enable_number <newline><^END>
enable_number is the sum of the binary-weighted values of the Enable register bits.
The return format can be selected by the
:FORMat:SREGister command. Response
data type is NR1 or NDN.
Remarks Bit definitions of the Standard Event register are shown in Table 3-2. All of the
enabled events of the Standard Event Status Enable register are logically ORed to
cause the Event Summary Bit (ESB) of the Status Byte register to be set.
The *CLS (clear status) command will not clear the enable register but it does clear
all bits in the event register.
The :STATus:PRESet command does not clear the bits in the Status Byte register.
See Also :FORMat:SREGister and :STATus:PRESet

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals