4- 132 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
Syntax [:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO mode
[:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables automatic ranging. If this function is disabled, the source
output is performed by using the range set by the
[:SOURce]:<CURRent|VOLTage>:RANGe command.
mode=1 or ON enables automatic ranging. If this function is enabled, the channel
automatically sets the range which provides the best resolution to apply the source
output.
If a range is manually selected, automatic ranging is disabled.
Query response mode <newline>
mode is 0 or 1, and indicates that automatic ranging is off or on, respectively.
Response data type is NR1.
Example :CURR:RANG:AUTO 0
:SOUR2:VOLT:RANG:AUTO?
[:SOURce]:<CURRent|VOLTage>:RANGe:AUTO:LLI
Mit
Specifies the lower limit for the automatic output ranging operation, and sets the
minimum range which provides the best resolution to apply the specified value.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO:LLIMit range
[:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO:LLIMit? [range]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter range value (see “Source Output Ranges” on page 2-36) |MINimum|
MAXimum|DEFault. Parameter data type is NRf+. Query does
not support range=value.
value for current output: See Table 2-6.