Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 197
Subsystem Commands
TRIGger Subsystem
:TRIG2:TRAN:BYP?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:COUNt
Sets the trigger count for the specified device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:COUNt trigger_count
:TRIGger[c]<:ACQuire|:TRANsient>:COUNt? [trigger_count]
:TRIGger[c][:ALL]:COUNt? trigger_count
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter trigger_count Trigger count. value (1 to 100000 or 2147483647)|INFinity|
MINimum|MAXimum|DEFault (default is 1). Parameter data
type is NRf+. value=2147483647 indicates infinity.
Query does not support trigger_count=value and INFinity.
Arm count × Trigger count must be less than 100001.
Query response response <newline>
response returns the present setting of trigger count. If a parameter is specified,
response returns the value assigned to DEF, MIN, or MAX. Response data type is
NR1.
Example :TRIG:COUN 10
:TRIG2:TRAN:COUN?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:DELay
Sets the trigger delay for the specified device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:DELay delay
:TRIGger[c]<:ACQuire|:TRANsient>:DELay? [delay]
:TRIGger[c][:ALL]:DELay? delay
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.