Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 105
Subsystem Commands
SOURce Subsystem
Parameter data Ramp waveform end level. value (see “Source Output Ranges”
on page 2-36) |MINimum|MAXimum|DEFault (default is 0.0).
Parameter data type is NRf+. Query does not support
data=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:RAMP:END 1E-3
:SOUR2:ARB:CURR:RAMP:END?
[:SOURce]:ARB:<CURRent|VOLTage>:RAMP:END:T
IME
Sets the end time of the current or voltage ramp waveform output. Waveform period
must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:END:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:END:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter time Ramp waveform end time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:RAMP:END:TIME 0.1
:SOUR2:ARB:CURR:RAMP:END:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:RAMP:RTIMe
Sets the rise time of the current or voltage ramp waveform output. Waveform period
must not exceed 1000 seconds.