Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TRIangle:STARt[:LEVel] data
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TRIangle:STARt[:LEVel]? [data]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter data Triangle waveform start level. value (see “Source Output
Ranges” on page 2-36) |MINimum|MAXimum|DEFault
(default is 0.0). Parameter data type is NRf+. Query does not
support data=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:TRI:STAR 1E-3
:SOUR2:ARB:CURR:TRI:STAR?
[:SOURce]:ARB:<CURRent|VOLTage>:TRIangle:STA
Rt:TIME
Sets the start time of the current or voltage triangle waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TRIangle:STARt:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TRAPezoid:STARt:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter time Triangle waveform start time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:TRI:STAR:TIME 0.1
:SOUR2:ARB:CURR:TRI:STAR:TIME?