Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 119
Subsystem Commands
SOURce Subsystem
Parameter fall_time Triangle waveform fall time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support fall_time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:TRI:FTIM 0.1
:SOUR2:ARB:CURR:TRI:FTIM?
[:SOURce]:ARB:<CURRent|VOLTage>:TRIangle:RTI
Me
Sets the rise time of the current or voltage triangle waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TRIangle:RTIMe rise_time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TRIangle:RTIMe? [rise_time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter rise_time Triangle waveform rise time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support rise_time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:TRI:RTIM 0.1
:SOUR2:ARB:CURR:TRI:RTIM?
[:SOURce]:ARB:<CURRent|VOLTage>:TRIangle:STA
Rt[:LEVel]
Sets the start level of the current or voltage triangle waveform output.