EasyManua.ls Logo

Keysight Technologies B2961A - :DISPlay[:WINDow]:TEXT:STATe; :DISPlay:ZOOM

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 13
Subsystem Commands
DISPlay Subsystem
:DISPlay[:WINDow]:TEXT:STATe
Shows or hides the text message set by the :DISPlay[:WINDow]:TEXT:DATA
command.
Syntax :DISPlay[:WINDow[d]]:TEXT:STATe mode
:DISPlay[:WINDow[d]]:TEXT:STATe?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=1 or ON shows the text message.
mode=0 or OFF hides the text message.
Query response mode <newline>
mode is 0 or 1, and indicates that the text message is off or on, respectively.
Response data type is NR1.
Example :DISP:TEXT:STAT 1
:DISPlay:ZOOM
Enables or disables the zoom function of the front panel display. This function is
effective for the Dual view and Single view.
If this function is enabled, the setup information is not displayed and the
measurement result is zoomed. Then,
the Dual view displays the primary measurement data with a large font, and the
secondary measurement data with a small font, for each channel.
the Single view displays both primary and secondary measurement data with a
large font.
Syntax :DISPlay:ZOOM mode
:DISPlay:ZOOM?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=1 or ON enables the zoom function of the front panel display.
mode=0 or OFF disables the zoom function of the front panel display.
Query response mode <newline>