4- 194 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
TRIGger Subsystem
Example :ARM:TIM 2E-4
:ARM2:TRAN:TIM?
:ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:TOUT
put:SIGNal
Selects the trigger output for the status change between the idle state and the arm
layer. Multiple trigger output ports can be set.
Syntax :ARM[c]<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:TOUTput:SIGNal
output{,output}
:ARM[c]<:ACQuire|:TRANsient>[:LAYer]:TOUTput:SIGNal?
For <:ACQuire|:TRANsient|[:ALL]> and <:ACQuire|:TRANsient>, specify
:ACQuire for measurement, :TRANsient for source output, or :ALL for both device
actions.
Parameter output Trigger output port. EXT1 (default)|EXT2|EXT3|EXT4|
EXT5|EXT6|EXT7|EXT8|EXT9|EXT10|EXT11|EXT12|
EXT13|EXT14|LAN|INT1|INT2. Parameter data type is CPD.
output=INT1 or INT2 selects the internal bus 1 or 2, respectively.
output=LAN selects a LAN port.
output=EXTn selects the GPIO pin n, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting, INT1, INT2, LAN, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :ARM:TOUT:SIGN EXT1
:ARM2:TRAN:TOUT:SIGN?
:ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:TOUT
put[:STATe]
Enables or disables the trigger output for the status change between the idle state
and the arm layer.