4- 182 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SYSTem Subsystem
Example :SYST:TIME:TIM:COUN?
:SYSTem:TIME:TIMer:COUNt:RESet:AUTO
Enables or disables the automatic reset function of the timer. If this function is
enabled, the timer count is reset when the initiate action occurs.
Syntax :SYSTem:TIME:TIMer:COUNt:RESet:AUTO mode
:SYSTem:TIME:TIMer:COUNt:RESet:AUTO?
Parameter mode Automatic reset function on or off. 0|OFF|1|ON (default).
Parameter data type is boolean.
mode=1 or ON enables the automatic reset function.
mode=0 or OFF disables the automatic reset function.
Query response mode <newline>
mode returns 0 or 1, and indicates that the automatic reset function is off or on,
respectively. Response data type is NR1.
Example :SYST:TIME:TIM:COUN:RES:AUTO 0
:SYST:TIME:TIM:COUN:RES:AUTO?
:SYSTem:TIME:TIMer:COUNt:RESet[:IMMediate]
Resets the timer count immediately.
Syntax :SYSTem:TIME:TIMer:COUNt:RESet[:IMMediate]
Example :SYST:TIME:TIM:COUN:RES