4- 68 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
OUTPut Subsystem
Syntax :OUTPut[c]:PROTection[:STATe] mode
:OUTPut[c]:PROTection[:STATe]?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables the over voltage/current protection.
mode=1 or ON enables the over voltage/current protection.
Query response mode <newline>
mode is 0 or 1, and indicates that the over voltage/current protection is off or on,
respectively. Response data type is NR1.
Example :OUTP:PROT 1
:OUTP2:PROT:STAT?
:OUTPut:RECall
Recalls the channel setup saved by the :OUTPut:SAVE command.
Syntax :OUTPut[c]:RECall index
Parameter index 0|1. Parameter data type is NR1.
index=0 is used to recall channel setup 0.
index=1 is used to recall channel setup 1.
Example :OUTP:REC 1
:OUTPut:RESistance:EMULate:<CURRent|VOLTage>
Sets the current or voltage list used for performing the emulation mode operation of
the programmable output resistance function. Both the current list and the voltage
list must be set to decide the output resistance. Also the source type list must be set
by the
:OUTPut:RESistance:EMULate:MODE command to specify the channel
operation, voltage source or current source.
The order of data set to the list is important. The current value, voltage value, and
source type are read in the listed order and used for performing the emulation mode
operation of the programmable output resistance function.