4- 138 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
type=BEFore enables trigger output at the beginning of arm, trigger, and device
actions (transient or acquire).
type=AFTer enables trigger output at the end of arm, trigger, and device actions
(transient or acquire).
type=BOTH enables trigger output at both beginning and end of arm, trigger, and
device actions (transient or acquire).
Query response response <newline>
response returns the present setting of output trigger timing, BEF, AFT or BOTH.
Response data type is CRD.
Example :DIG:EXT:TOUT:POS BEF
:SOUR:DIG:EXT2:TOUT:POS?
[:SOURce]:DIGital:EXTernal:TOUTput[:EDGE]:WID
Th
Sets the pulse width of the output trigger for the specified GPIO pin.
Syntax [:SOURce]:DIGital:EXTernal[n]:TOUTput[:EDGE]:WIDTh width
[:SOURce]:DIGital:EXTernal[n]:TOUTput[:EDGE]:WIDTh? [width]
Parameter width Pulse width. value (1E-5 to 1E-2, in seconds)|
MINimum|MAXimum|DEFault (default is 0.1 ms). Parameter
data type is NRf+. Query does not support width=value.
Query response width <newline>
width returns the present setting. If a parameter is specified, width returns the value
assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :DIG:EXT:TOUT:WIDT 1E-5
:SOUR:DIG:EXT14:TOUT:WIDT?
[:SOURce]:DIGital:EXTernal:TOUTput:TYPE
Selects the output trigger type for the specified GPIO pin.
Syntax [:SOURce]:DIGital:EXTernal[n]:TOUTput:TYPE type