EasyManuals Logo

Keysight Technologies B2961A Command Reference

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
3- 12 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Common Commands
*SAV
*SAV
This command stores the present state of the instrument to the specified location in
non-volatile memory. Up to 10 states can be stored in the memory locations 0
through 9. Any state previously stored in the same location will be overwritten. Use
the
*RCL command to retrieve instrument states.
Syntax *SAV memory
Parameter memory One of the memory locations 0 to 9. Parameter data type is NR1.
The locations 0 to 4 are in the non-volatile memory, and the
locations 5 to 9 are in the volatile memory.
Remarks If a particular state is desired at power-on, it should be stored in the location 0. It
will then be automatically recalled at power turn-on if the Output Power-On state is
set to *RCL 0.
Data described in “Non-Volatile Settings” on page 1-21 is not affected by the *SAV
command.
The *RST command does not affect the configurations stored in memory. Once a
state is stored, it remains until it is overwritten or specifically deleted.
CAUTION This command causes a write cycle to non-volatile memory. Non-volatile memory
has a finite maximum number of write cycles. Programs that repeatedly cause write
cycles to non-volatile memory can eventually exceed the maximum number of write
cycles and cause the memory to fail.

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals