Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 109
Subsystem Commands
SOURce Subsystem
Parameter offset Offset level of sinusoidal waveform. value (see “Source Output
Ranges” on page 2-36) |MINimum|MAXimum|DEFault
(default is 0.0). Parameter data type is NRf+. Query does not
support offset=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SIN:OFFS 1E-3
:SOUR2:ARB:CURR:SIN:OFFS?
[:SOURce]:ARB:<CURRent|VOLTage>:SINusoid:PMA
Rker[:PHASe]
Specifies the position of the phase marker for the current or voltage sinusoidal
waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker[:PHASe] phase
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker[:PHASe]? [phase]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter phase Phase, in degree. value (0 to 360)|MINimum|MAXimum|
DEFault (default is 0). Parameter data type is NRf+. Query does
not support phase=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SIN:PMAR:PHAS 180
:SOUR2:ARB:CURR:SIN:PMAR:PHAS?