EasyManuals Logo

Keysight Technologies B2961A Command Reference

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #336 background imageLoading...
Page #336 background image
4- 198 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
TRIGger Subsystem
Parameter delay Trigger delay, in seconds. value (0 to 100000)|MINimum|
MAXimum|DEFault (default is 0). Parameter data type is
NRf+. Query does not support delay=value.
Query response response <newline>
response returns the present setting of trigger delay. If a parameter is specified,
response returns the value assigned to DEF, MIN, or MAX. Response data type is
NR3.
Example :TRIG:DEL 0.1
:TRIG2:TRAN:DEL?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>[:IMMediate
]
Sends an immediate trigger for the specified device action to the specified channel.
When the status of the specified device action is initiated, the trigger causes the
specified device action.
Syntax :TRIGger<:ACQuire|:TRANsient|[:ALL]>[:IMMediate] [chanlist]
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter chanlist Channels. Parameter data type is channel list.
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See
“Channel List
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1),
and (@2:1) selects both channels 1 and 2.
If this parameter is not specified, chanlist=(@1) is set.
Example :TRIG:ACQ (@2)
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:SOURce:L
AN
Specifies one or more LXI triggers used for the trigger source for the specified
device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:SOURce:LAN lan_id{,lan_id}

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals