EasyManua.ls Logo

Keysight Technologies B2961A - :ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOUR ce:LAN; :ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOUR ce[:SIGNal]

Keysight Technologies B2961A
364 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4- 192 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
TRIGger Subsystem
:ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOUR
ce:LAN
Specifies one or more LXI triggers used for the arm source for the specified device
action.
Syntax :ARM[c]<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOURce:LAN lan_id{,lan_id}
:ARM[c]<:ACQuire|:TRANsient>[:LAYer]:SOURce:LAN?
For <:ACQuire|:TRANsient|[:ALL]> and <:ACQuire|:TRANsient>, specify
:ACQuire for measurement, :TRANsient for source output, or :ALL for both device
actions.
Parameter lan_id LAN ID of the LXI trigger. LAN0|LAN1|LAN2|LAN3|LAN4|
LAN5|LAN6|LAN7. All is selected as default. Parameter data
type is CPD.
Query response response <newline>
response returns the present setting, LAN0 through LAN7. Response data type is
CRD. Multiple responses are separated by a comma.
Example :ARM:SOUR:LAN LAN7
:ARM2:TRAN:SOUR:LAN?
:ARM<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOUR
ce[:SIGNal]
Selects the arm source for the specified device action.
Syntax :ARM[c]<:ACQuire|:TRANsient|[:ALL]>[:LAYer]:SOURce[:SIGNal] source
:ARM[c]<:ACQuire|:TRANsient>[:LAYer]:SOURce[:SIGNal]?
For <:ACQuire|:TRANsient|[:ALL]> and <:ACQuire|:TRANsient>, specify
:ACQuire for measurement, :TRANsient for source output, or :ALL for both device
actions.
Parameter source Arm source. AINT (default)|BUS|TIMer|INT1|INT2|LAN|
EXT1|EXT2|EXT3|EXT4|EXT5|EXT6|EXT7|EXT8|EXT9|EX
T10| EXT11|EXT12| EXT13|EXT14. Parameter data type is
CPD.