Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 65
Subsystem Commands
OUTPut Subsystem
Query response mode <newline>
mode is 0 or 1, and indicates that the high capacitance mode is off or on,
respectively. Response data type is NR1.
Example :OUTP:HCAP 1
:OUTP2:HCAP:STAT?
:OUTPut:LOW
Selects the state of the low terminal. Before executing this command, the source
output must be disabled by the
:OUTPut[:STATe] command. Or else, an error
occurs.
Syntax :OUTPut[c]:LOW low_state
:OUTPut[c]:LOW?
Parameter low_state FLOat|GROund (default). Parameter data type is CPD.
low_state=FLOat sets the floating state.
low_state=GROund sets the ground state. The low terminal is connected to ground.
Query response low_state <newline>
low_state is FLO or GRO, and indicates the low terminal state. Response data type
is CRD.
Example :OUTP:LOW FLO
:OUTP2:LOW?
:OUTPut:OFF:AUTO
Enables or disables the automatic output off function.
Syntax :OUTPut[c]:OFF:AUTO mode
:OUTPut[c]:OFF:AUTO?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables the automatic output off function.