Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 189
Subsystem Commands
TRIGger Subsystem
TRIGger Subsystem
For the numeric suffix [c], see “Numeric Suffix” on page 1-8.
:ABORt<:ACQuire|:TRANsient|[:ALL]>
Aborts the specified device action for the specified channel. Trigger status is
changed to idle.
Syntax :ABORt<:ACQuire|:TRANsient|[:ALL]> [chanlist]
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter chanlist Channels. Parameter data type is channel list.
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See
“Channel List
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1),
and (@2:1) selects both channels 1 and 2.
If this parameter is not specified, chanlist=(@1) is set.
Example :ABOR:ACQ (@2)
:ARM<:ACQuire|:TRANsient|[:ALL]>[:IMMediate]
Sends an immediate arm trigger for the specified device action to the specified
channel. When the status of the specified device action is initiated, the arm trigger
causes a layer change from arm to trigger.
Syntax :ARM<:ACQuire|:TRANsient|[:ALL]>[:IMMediate] [chanlist]
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter chanlist Channels. Parameter data type is channel list.
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See
“Channel List
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1),
and (@2:1) selects both channels 1 and 2.