Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 103
Subsystem Commands
SOURce Subsystem
Waveform period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:EXPonent:STARt:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:EXPonent:STARt:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter time Exponential waveform start time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:EXP:STAR:TIME 0.1
:SOUR2:ARB:CURR:EXP:STAR:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:EXPonent:TC
ONstant
Sets the time constant of the current or voltage exponential waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TCONstant time_constant
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TCONstant? [time_constant]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter time_constant Exponential waveform time constant, in seconds. value (0 to
1000)|MINimum|MAXimum|DEFault (default is 0). Parameter
data type is NRf+. Query does not support
time_constant=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:EXP:TCON 0.1