EasyManuals Logo

Keysight Technologies B2961A Command Reference

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 59
Subsystem Command Summary
Reading Source/Measure Data
Table 2-12 FORMat Subsystem
Command Summary Reset setting
:FORM:BORD byte_order
:FORM:BORD?
Sets the byte order of binary output data.
byte_order=NORMal|SWAPped
NORM
:FORM[:DATA] format
:FORM[:DATA]?
Sets the data output format.
format=ASCii|REAL,32|REAL,64
ASC
:FORM:DIG format
:FORM:DIG?
Sets the response format of the bit
pattern.
format=ASCii|BINary|OCTal|
HEXadecimal
ASC
:FORM:ELEM:CALC type{,type}
:FORM:ELEM:CALC?
Specifies the elements included in the
calculation result data.
type=CALC|TIME|STATus
Order of returned data: calc, time, status
CALC
:FORM:ELEM:SENS type{,type}
:FORM:ELEM:SENS?
Specifies the elements included in the
sense or measurement result data.
type=VOLTage|CURRent|RESistance|
TIME|STATus|SOURce
Order of returned data: voltage, current,
resistance, time, status, source
VOLT,
CURR, RES,
TIME, STAT,
SOUR
:FORM:SREG format
:FORM:SREG?
Sets the response format of the status byte
register.
format=ASCii|BINary|OCTal|
HEXadecimal
ASC

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals