Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 59
Subsystem Command Summary
Reading Source/Measure Data
Table 2-12 FORMat Subsystem
Command Summary Reset setting
:FORM:BORD byte_order
:FORM:BORD?
Sets the byte order of binary output data.
byte_order=NORMal|SWAPped
NORM
:FORM[:DATA] format
:FORM[:DATA]?
Sets the data output format.
format=ASCii|REAL,32|REAL,64
ASC
:FORM:DIG format
:FORM:DIG?
Sets the response format of the bit
pattern.
format=ASCii|BINary|OCTal|
HEXadecimal
ASC
:FORM:ELEM:CALC type{,type}
:FORM:ELEM:CALC?
Specifies the elements included in the
calculation result data.
type=CALC|TIME|STATus
Order of returned data: calc, time, status
CALC
:FORM:ELEM:SENS type{,type}
:FORM:ELEM:SENS?
Specifies the elements included in the
sense or measurement result data.
type=VOLTage|CURRent|RESistance|
TIME|STATus|SOURce
Order of returned data: voltage, current,
resistance, time, status, source
VOLT,
CURR, RES,
TIME, STAT,
SOUR
:FORM:SREG format
:FORM:SREG?
Sets the response format of the status byte
register.
format=ASCii|BINary|OCTal|
HEXadecimal
ASC