5- 8 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Error Messages
Execution Error
Execution Error
Keysight B2961A/B2962A reports -2XX errors when it is unable to perform a valid
programming command.
Error -200 Execution error
Generic execution error for B2961A/B2962A that cannot be determined more
specifically.
Error -220 Parameter error; message
Invalid parameter was specified. Set appropriate value.
Error -221 Settings conflict; message; channel n
A specified parameter setting could not be executed due to the present device state.
Check the settings specified by message and channel n, and set appropriate value.
Error -222 Data out of range; message; channel n
Interpreted value of the program was out of range as defined by B2961A/B2962A.
Check the B2961A/B2962A settings specified by message and channel n, and set
appropriate value.
Error -223 Too much data
Too many parameters were sent. Reduce number of list data.
Error -224 Illegal parameter value; message
Illegal parameter value was sent. Set appropriate parameter value.
Error -230 Data corrupt or stale
Possibly invalid data; new reading started but not completed since last access.
Error -231 Data questionable
Measurement accuracy is suspect.
Error -232 Invalid format
The data format or structure is inappropriate.

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals