4- 106 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:RTIMe rise_time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:RTIMe? [rise_time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter rise_time Ramp waveform rise time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support rise_time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:RAMP:RTIM 0.1
:SOUR2:ARB:CURR:RAMP:RTIM?
[:SOURce]:ARB:<CURRent|VOLTage>:RAMP:STARt
[:LEVel]
Sets the start level of the current or voltage ramp waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:STARt[:LEVel] data
[:SOURce[c]]:ARB:<CURRent|VOLTage>:RAMP:STARt[:LEVel]? [data]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter data Ramp waveform start level. value (see “Source Output Ranges”
on page 2-36) |MINimum|MAXimum|DEFault (default is 0.0).
Parameter data type is NRf+. Query does not support
data=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:RAMP:STAR 1E-3
:SOUR2:ARB:CURR:RAMP:STAR?