4- 108 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SIN:AMPL 1E-3
:SOUR2:ARB:CURR:SIN:AMPL?
[:SOURce]:ARB:<CURRent|VOLTage>:SINusoid:FRE
Quency
Sets the frequency of the current or voltage sinusoidal waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:FREQuency frequency
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:FREQuency? [frequency]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter frequency Frequency of sinusoidal waveform, in Hz. value (1.0E-3 to
1.0E+4)|MINimum|MAXimum|DEFault (default is 1.0E-3).
Parameter data type is NRf+. Query does not support
frequency=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SIN:FREQ 0.1
:SOUR2:ARB:CURR:SIN:FREQ?
[:SOURce]:ARB:<CURRent|VOLTage>:SINusoid:OFF
Set
Sets the offset level of the current or voltage sinusoidal waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:OFFSet offset
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:OFFSet? [offset]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.