4- 110 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
[:SOURce]:ARB:<CURRent|VOLTage>:SINusoid:PMA
Rker:SIGNal
Selects the trigger output at the phase marker of the current or voltage sinusoidal
waveform output. Multiple trigger output ports can be set.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker:SIGNal
output{,output}
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker:SIGNal?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter output Trigger output port. EXT1 (default)|EXT2|EXT3|EXT4|EXT5|
EXT6|EXT7|EXT8|EXT9|EXT10|EXT11|EXT12|EXT13|
EXT14|INT1|INT2. Parameter data type is CPD.
output=INT1 or INT2 selects the internal bus 1 or 2, respectively.
output=EXTn selects the GPIO pin n, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting, INT1, INT2, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :ARB:CURR:SIN:PMAR:SIGN EXT3
:SOUR2:ARB:CURR:SIN:PMAR:SIGN?
[:SOURce]:ARB:<CURRent|VOLTage>:SINusoid:PMA
Rker:STATe
Enables or disables the trigger output at the phase marker of the current or voltage
sinusoidal waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker:STATe mode
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SINusoid:PMARker:STATe?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.