4- 114 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SOURce Subsystem
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SQU:TOP:TIME 0.1
:SOUR2:ARB:CURR:SQU:TOP:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:TRAPezoid:E
ND:TIME
Sets the end time of the current or voltage trapezoidal waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TRAPezoid:END:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TRAPezoid:END:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.
Parameter time Trapezoidal waveform end time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:TRAP:END:TIME 0.1
:SOUR2:ARB:CURR:TRAP:END:TIME?
[:SOURce]:ARB:<CURRent|VOLTage>:TRAPezoid:FT
IMe
Sets the fall time of the current or voltage trapezoidal waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:TRAPezoid:FTIMe fall_time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:TRAPezoid:FTIMe? [fall_time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage fo r
voltage output.