4- 160 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
STATus Subsystem
:STATus:QUEStionable:<CALibration|CURRent|TEM
Perature|TEST|VOLTage>:ENABle
Sets the questionable calibration, current, temperature, test, or voltage status enable
register. The enable register is a mask which allows true conditions in the event
register to be reported in the summary bit.
Syntax :STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:E
NABle mask
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:E
NABle?
For <CALibration|CURRent|TEMPerature|TEST|VOLTage>, specify CALibration
for the questionable calibration status condition register, CURRent for the
questionable current status condition register, TEMPerature for the questionable
temperature status condition register, TEST for the questionable self-test status
condition register, or VOLTage for the questionable voltage status condition register.
Parameter mask Mask. 0 to 65535 (decimal). Default is 0. Parameter data type is
NR1 or NDN.
mask is the sum of the binary-weighted values for the set bits.
Query response mask <newline>
mask returns the present setting of the specified enable register. Response data type
is NR1 (decimal) or NDN (binary, octal, or hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:QUES:CURR:ENAB 65535
:STAT:QUES:TEMP:ENAB?
:STATus:QUEStionable:<CALibration|CURRent|TEM
Perature|TEST|VOLTage>[:EVENt]?
Returns the value of the questionable calibration, current, temperature, test, or
voltage status event register. The register setting is changed by this command.
Syntax :STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>[:
EVENt]?