4- 66 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
OUTPut Subsystem
mode=1 or ON enables the automatic output off function. If this function is enabled,
the source output is automatically turned off immediately when the grouped
channels change status from busy to idle.
Query response mode <newline>
mode is 0 or 1, and indicates that the automatic output off function is off or on,
respectively. Response data type is NR1.
Example :OUTP:OFF:AUTO 1
:OUTP2:OFF:AUTO?
:OUTPut:OFF:MODE
Selects the source condition after output off.
Syntax :OUTPut[c]:OFF:MODE mode
:OUTPut[c]:OFF:MODE?
Parameter mode ZERO|HIZ|NORMal (default). Parameter data type is CPD.
mode=NORMal selects the following setup.
Source function: Voltage source
Output voltage: 0 V
Current compliance: 100 μA at the 100 μA range
Output relay: off (open or break)
mode=HIZ selects the following setup.
Output relay: off (open or break)
Voltage source setup is not changed if the source applies 40 V or less.
Current source setup is not changed if the source uses the 100 mA range or
lower.
mode=ZERO selects the following setup.
Source function: Voltage source
Output voltage: 0 V
Current compliance: 100 μA at the 100 μA range