Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 61
Subsystem Commands
OUTPut Subsystem
OUTPut Subsystem
For the numeric suffix [c], see “Numeric Suffix” on page 1-8.
:OUTPut:FILTer:EXTernal:TYPE
Selects the external filter type.
Syntax :OUTPut[c]:FILTer:EXTernal:TYPE type
:OUTPut[c]:FILTer:EXTernal:TYPE?
Parameter type LNF|ULNF (default)|HCULNF. Parameter data type is CPD.
type=LNF selects the low noise filter.
type=ULNF selects the ultra low noise filter.
type=HCULNF selects the high current ultra low noise filter.
Query response type <newline>
type is LNF or ULNF. Response data type is CRD.
Example :OUTP:FILT:EXT:TYPE LNF
:OUTP2:FILT:EXT:TYPE?
:OUTPut:FILTer:EXTernal:STATe
Enables or disables the external filter.
Syntax :OUTPut[c]:FILTer:EXTernal:STATe mode
:OUTPut[c]:FILTer:EXTernal:STATe mode?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables the external filter.
mode=1 or ON enables the external filter.
Query response mode <newline>