Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 173
Subsystem Commands
SYSTem Subsystem
code returns the error/event code. Multiple responses are listed in the FIFO
(first-in-first-out) order, separated by a comma. Response data type is NR1.
If the queue is empty, the response is +0.
Example :SYST:ERR:CODE:ALL?
:SYSTem:ERRor:CODE[:NEXT]?
Reads and removes the top item in the error/event queue, and returns the top code.
Syntax :SYSTem:ERRor:CODE[:NEXT]?
Query response code <newline>
code returns the error/event code. Response data type is NR1.
If the queue is empty, the response is +0.
Example :SYST:ERR:CODE?
:SYSTem:ERRor:COUNt?
Returns the number of items in the error/event queue.
Syntax :SYSTem:ERRor:COUNt?
Query response response <newline>
response returns the number of items. Response data type is NR1.
If the queue is empty, the response is +0.
Example :SYST:ERR:COUN?
:SYSTem:ERRor[:NEXT]?
Reads and removes the top item in the error/event queue, and returns the top code
and message.
Syntax :SYSTem:ERRor[:NEXT]?
Query response response <newline>