Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 133
Subsystem Commands
SOURce Subsystem
value for voltage output: See Table 2-4.
Query response range <newline>
range returns the present setting. If a parameter is specified, range returns the value
assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :CURR:RANG:AUTO:LLIM 1E-6
:SOUR2:VOLT:RANG:AUTO:LLIM?
[:SOURce]:<CURRent|VOLTage>:RANGe:RPRiority
Selects the output pulse priority mode, low noise priority or transient speed priority,
of the specified source channel.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:RANGe:RPRiority mode
[:SOURce[c]]:<CURRent|VOLTage>:RANGe:RPRiority?
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter mode NOISe (default)|TRANsient. Parameter data type is CPD.
mode=NOISe sets the low noise priority mode to obtain clean output pulse.
mode=TRANsient sets the transient speed priority mode to obtain high slew rate of
output pulse. This mode is effective when the 1 A, 1.5 A or 10 A range is used for
the current output or the current measurement.
Query response mode <newline>
mode returns NOIS or TRAN. Response data type is CRD.
Example :CURR:RANG:RPR TRAN
:SOUR2:VOLT:RANG:RPR?
[:SOURce]:<CURRent|VOLTage>:<STARt|STOP>
Sets the start or stop value for the current or voltage sweep output.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:<STARt|STOP> data
[:SOURce[c]]:<CURRent|VOLTage>:<STARt|STOP>? [data]