4- 90 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
SENSe Subsystem
:SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:PROTection[:LEVel]:NEGative?
[compliance]
For <CURRent[:DC]|VOLTage[:DC]>, specify CURRent[:DC] for current
compliance or VOLTage[:DC] for voltage compliance.
Parameter compliance value (see maximum current or maximum voltage in “Source
Output Ranges” on page 2-36) |MINimum| MAXimum|DEFault
(default is 100 μA or 2 V). Parameter data type is NRf+.
Effective values of value are from the minimum measurement
value to the maximum measurement value of the channel.
Query does not support compliance=value.
Query response compliance <newline>
compliance returns the present compliance value. If a parameter is specified,
compliance returns the value assigned to DEF, MIN, or MAX. Response data type is
NR3.
Example :SENS:CURR:PROT:NEG 1E-3
:SENS2:CURR:DC:PROT:NEG?
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:PROTection
[:LEVel]:POSitive
Sets the positive side compliance value of the specified channel.
Syntax :SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:PROTection[:LEVel]:POSitive
compliance
:SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:PROTection[:LEVel]:POSitive?
[compliance]
For <CURRent[:DC]|VOLTage[:DC]>, specify CURRent[:DC] for current
compliance or VOLTage[:DC] for voltage compliance.
Parameter compliance value (see maximum current or maximum voltage in “Source
Output Ranges” on page 2-36) |MINimum| MAXimum|DEFault
(default is 100 μA or 2 V). Parameter data type is NRf+.
Effective values of value are from the minimum measurement
value to the maximum measurement value of the channel.
Query does not support compliance=value.