EasyManuals Logo

Keysight Technologies B2961A Command Reference

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
3- 14 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Common Commands
*SRE
Table 3-3 Status Byte Register Bit Definitions
bit
decimal
value
description definition
0 1 Measurement status summary One or more bits are set in the Measurement Status
register (bits must be enabled, see
:STATus:<MEASurement|OPERation|QUEStionabl
e>:ENABle command).
1 2 Not used 0 is returned.
2 4 Error queue not empty One or more errors have been stored in the Error
Queue (see
:SYSTem:ERRor[:NEXT]? command).
3 8 Questionable status summary One or more bits are set in the Questionable Status
register (bits must be enabled, see
:STATus:<MEASurement|OPERation|QUEStionabl
e>:ENABle command).
4 16 Output buffer Data is available in the instrument’s output buffer.
5 32 Event status byte summary One or more bits are set in the Standard Event
register (bits must be enabled, see
*ESE command).
6 64 Master status summary
(Request for service)
One or more bits are set in the Status Byte register
(bits must be enabled, see
*SRE command). Also
used to indicate a request for service.
7 128 Operation status summary One or more bits are set in the Operation Status
register (bits must be enabled, see
:STATus:<MEASurement|OPERation|QUEStionabl
e>:ENABle command).

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals