4- 158 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
STATus Subsystem
Syntax :STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:C
ONDition?
For <CALibration|CURRent|TEMPerature|TEST|VOLTage>, specify CALibration
for the questionable calibration status condition register, CURRent for the
questionable current status condition register, TEMPerature for the questionable
temperature status condition register, TEST for the questionable self-test status
condition register, or VOLTage for the questionable voltage status condition register.
Query response value <newline>
value returns the value of the specified register. It is the sum of the binary-weighted
values for the set bits. Response data type is NR1 (decimal) or NDN (binary, octal,
or hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:QUES:CAL:COND?
:STAT:QUES:CURR:COND?
:STAT:QUES:TEMP:COND?
:STAT:QUES:TEST:COND?
:STAT:QUES:VOLT:COND?
Table 4-7 Questionable Calibration Register Bit Definitions
bit
decimal
value
description definition
0 1 Ch1 Calibration Failed Calibration fail in channel 1.
1 2 Ch2 Calibration Failed Calibration fail in channel 2.
3 to 15 Not used 0 is returned.

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals