Keysight B2961A/B2962A SCPI Command Reference, Edition 6 5- 17
Error Messages
B2961A/B2962A Specific Error
Internal communication failure was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 352 Emergency; Watchdog timer expired; Channel n
Watchdog timer expired status was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 353 Emergency; F-COM CPLD reset detected; Channel n
F-COM CPLD reset status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 354 Emergency; VADC data was lost; Channel n
Channel n voltage ADC data was lost. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 355 Emergency; IADC data was lost; Channel n
Channel n current ADC data was lost. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 356 Emergency; Sense data FIFO overflow detected; Channel n
Sense data FIFO overflow was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 361 Emergency; Internal communication failure detected by
CPU; Channel n
Channel n internal communication failure was detected by CPU. All channel output
is changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 362 Emergency; Internal command queue overflow detected;
Channel n
Internal command queue overflow was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 363 Emergency; Sense data was not received for acquire
trigger; Channel n
Channel n sense data was not received for acquire trigger. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.