Keysight B2961A/B2962A SCPI Command Reference, Edition 6 2- 25
Subsystem Command Summary
Setting Source Output and Measurement
[:SOUR[c]]:WAIT:GAIN gain
[:SOUR[c]]:WAIT:GAIN? [DEFault|
MINimum|MAXimum]
Sets the gain value used for calculating
the source wait time for the specified
channel.
gain=MINimum|MAXimum|DEFault|
0
to 100
1
[:SOUR[c]]:WAIT:OFFS offset
[:SOUR[c]]:WAIT:OFFS? [DEFault|
MINimum|MAXimum]
Sets the offset value used for calculating
the source wait time for the specified
channel.
offset=MINimum|MAXimum|DEFault|
0
to 1 seconds
0
[:SOUR[c]]:WAIT[:STAT] mode
[:SOUR[c]]:WAIT[:STAT]?
Enables or disables the source wait time
for the specified channel. This wait time
is defined as the time the source channel
cannot change the output after the start of
a DC output or the trailing edge of a
pulse.
mode=0 or OFF disables the source wait
time. The wait time is set to 0.
mode=1 or ON enables the source wait
time given by the following formula.
[:SOUR[c]]:WAIT:AUTO ON|1
condition:
wait time = gain × initial wait time +
offset
[:SOUR[c]]:WAIT:AUTO OFF|0
condition:
wait time = offset
The initial wait time is automatically set
by the instrument and cannot be changed.
ON