17.1 Touch Panel Calibration
17 - 1
17
MAINTENANCE
17. MAINTENANCE
17.1 Touch Panel Calibration
â– 1. Touch panel calibration setting function
Touch panel reading error can be corrected.
Normally the adjustment is not required, however, the difference between a touched position and the object position
may occur as the period of use elapses.
When any difference between a touched position and the object position occurs, correct the position with this function.
â– 2. Touch panel calibration setting display operation
Run Stop Run Stop
Before adjustment After adjustment
The [Run] will operate though you
intended to touch the [Stop] button.
The [Stop] button can be touched
without fail.
Main menu
Main menu
Calibration
Main menu
Touch
[Maintenance].
Touch
[Calibration].