Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 4-44 P/N 196515 (Revision A)
Go to Servo Controller
(PMAC) Verification
Procedure
Go to Interlock Fault
Isolation Procedure –
Main Power Verification
Does green
light on Beacon
illuminate?
Locate the power LEDs on
the XY Servo Interface PWA.
Go to Interlock Fault
Isolation Procedure –
Main Power Verification
Make sure Front Hatch
is closed.
Yes
No
Does red SRVO_
PWR LOW LED on XY
Servo Interface PWA
illuminate?
NoYes
Go to Servo Controller
(PMAC) Verification
Procedure
Go to Interlock Fault
Isolation Procedure –
Main Power Verification
Does green
light on Beacon
illuminate?
Does green
light on Beacon
illuminate?
Locate the power LEDs on
the XY Servo Interface PWA.
Go to Interlock Fault
Isolation Procedure –
Main Power Verification
Make sure Front Hatch
is closed.
Yes
No
Does red SRVO_
PWR LOW LED on XY
Servo Interface PWA
illuminate?
Does red SRVO_
PWR LOW LED on XY
Servo Interface PWA
illuminate?
NoYes
Diagram Sheet 4-4-3
XY-Axis Motion Fault Isolation Procedure - Interlock Verification

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nordson Asymtek X-1000 Series and is the answer not in the manual?

Nordson Asymtek X-1000 Series Specifications

General IconGeneral
BrandNordson
ModelAsymtek X-1000 Series
CategoryDispenser
LanguageEnglish

Related product manuals