Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 1-32 P/N 196515 (Revision A)
Information Sheet 1-7-3
Safety Precautions (Continued)
F. Additional Precautions for Service Personnel
1. Only qualified personnel should be permitted to perform maintenance and
troubleshooting procedures.
2. When performing service or maintenance work on the dispensing system in
areas marked with an Electrical Warning Label make sure to disconnect all
power cords and lock out/tag out equipment as detailed in “Lock Out/Tag
Out AC Power and Pneumatic Pressure” on page 1-12.
3. Before troubleshooting the Pneumatic Module, turn Main Air Regulator off (0
psi) or disconnect Main Air Supply.