Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 4-59 P/N 196515 (Revision A)
Go to Z Servo Amplifier
Fault Isolation Procedure
Go to “PMAC Watchdog
Error Verification PMAC
Initialization Fault Isolation
Procedure
Open FmNT by double-
clicking the FmNT icon.
Does the FmNT
message “30753 The
motor failed to close”
appear when
initializing FmNT?
Does the FmNT
message “30796 The
dispenser PMAC device
initialization has failed” appear
when initializing FmNT before
any other message?
The FmNT message “30416
Dispenser motor reference not
found. Select OK to home the
dispenser” should appear
during initialization process.
Watch how the Dispensing
Head homes (it should move
upward in the Z-axis, then
simultaneously move in the X
and Y-axis to the Home position
in the front left corner of the
dispensing chamber.
Does Dispensing
Head complete
homing process with
no FmNT error
messages?
From the FmNT Main Menu,
click on the Jog icon.
Did Dispensing Head
hit mechanical hard
stop in X or Y-axis
during homing
process?
In the Jog Window, point and
click at several locations in the
blue square to move the
Dispensing Head throughout
the dispensing chamber.
Did FmNT message “30735
The motors have gone open loop.
Press OK to continue and
reinitialize the motion
controller” appear?
Stop. No Z Motion Fault.
If fault reoccurs, contact
Asymtek.
Go to “PMAC Fatal Following
Error Verification in PMAC
Initialization Fault Isolation
Procedure
Go to PMAC Home
Sensor Fault Isolation
Procedure
Exit FmNT, if open
Home the Dispensing Head by
clicking on the OK button in the
bottom of Message Box, or
press [Enter] on the Keyboard.
No
Yes
No
Yes
No
Yes
No
Yes
NoYes
Go to Z Servo Amplifier
Fault Isolation Procedure
Go to “PMAC Watchdog
Error Verification” PMAC
Initialization Fault Isolation
Procedure
Open FmNT by double-
clicking the FmNT icon.
Does the FmNT
message “30753 The
motor failed to close”
appear when
initializing FmNT?
Does the FmNT
message “30753 The
motor failed to close”
appear when
initializing FmNT?
Does the FmNT
message “30796 The
dispenser PMAC device
initialization has failed” appear
when initializing FmNT before
any other message?
Does the FmNT
message “30796 The
dispenser PMAC device
initialization has failed” appear
when initializing FmNT before
any other message?
The FmNT message “30416
Dispenser motor reference not
found. Select OK to home the
dispenser” should appear
during initialization process.
Watch how the Dispensing
Head homes (it should move
upward in the Z-axis, then
simultaneously move in the X
and Y-axis to the Home position
in the front left corner of the
dispensing chamber.
Does Dispensing
Head complete
homing process with
no FmNT error
messages?
Does Dispensing
Head complete
homing process with
no FmNT error
messages?
From the FmNT Main Menu,
click on the Jog icon.
Did Dispensing Head
hit mechanical hard
stop in X or Y-axis
during homing
process?
Did Dispensing Head
hit mechanical hard
stop in X or Y-axis
during homing
process?
In the Jog Window, point and
click at several locations in the
blue square to move the
Dispensing Head throughout
the dispensing chamber.
Did FmNT message “30735
The motors have gone open loop.
Press OK to continue and
reinitialize the motion
controller” appear?
Did FmNT message “30735
The motors have gone open loop.
Press OK to continue and
reinitialize the motion
controller” appear?
Stop. No Z Motion Fault.
If fault reoccurs, contact
Asymtek.
Go to “PMAC Fatal Following
Error Verification” in PMAC
Initialization Fault Isolation
Procedure
Go to PMAC Home
Sensor Fault Isolation
Procedure
Exit FmNT, if open
Home the Dispensing Head by
clicking on the OK button in the
bottom of Message Box, or
press [Enter] on the Keyboard.
No
Yes
No
Yes
No
Yes
No
Yes
NoYes
Diagram Sheet 4-5-4
No Z-Axis Motion Fault Isolation Procedure - Servo Controller (PMAC) Verification

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nordson Asymtek X-1000 Series and is the answer not in the manual?

Nordson Asymtek X-1000 Series Specifications

General IconGeneral
BrandNordson
ModelAsymtek X-1000 Series
CategoryDispenser
LanguageEnglish

Related product manuals