Instrument Function
R&S
®
SMB100A
183Operating Manual 1407.0806.32 ─ 21
Remote command:
[:SOURce<hw>]:POWer:MODE on page 382
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Sets an automatically repeated sweep cycle.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR AUTO
SOUR:POW:MODE SWE
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute
Single Sweep" button, or by means remote trigger commands, e.g.
*TRG.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:POW:EXEC
RF Block