Instrument Function
R&S
®
SMB100A
130Operating Manual 1407.0806.32 ─ 21
If an instrument setting in which a sweep was activated is stored, the sweep is started
when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also loaded.
If the list has been deleted in the meantime, an error message appears when the
instrument setting is loaded. If the list has been overwritten in the meantime, the new
entries will be used.
Remote command:
:MMEMory:LOAD:STATe on page 309
*RCL on page 282
Recall Immediate x
Loads the selected configuration from one of the three intermediate memories.
If an instrument setting in which a sweep was activated is stored, the sweep is started
when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also loaded.
If the list has been deleted in the meantime, an error message appears when the
instrument setting is loaded. If the list has been overwritten in the meantime, the new
entries will be used.
A message appears if no instrument configuration is stored in this memory.
Remote command:
*RCL on page 282
File Manager
Accesses the "File Manager" dialog, see Chapter 5.2.8.4, "File Manager",
on page 130.
Remote command:
n.a.
5.2.8.4 File Manager
The "File Manager" is a tool very similar to a standard Windows Explorer and helps
you manage mass storage media and files stored on the R&S SMB.
You can perform the following tasks:
●
Copying multiple files from disk to other media and vice versa, see Chapter 5.2.8.5,
"Accessing the File System of the Instrument and Transferring Files from and to
the Instrument", on page 132
●
Copying files into another directory, see Copy and Paste
●
Renaming and deleting files, see Rename and Delete
●
Creating new directories on the follwoing drives:
– hard disk
– internal flash memory
– memory stick
See Create New Directory