Remote Control Commands
R&S
®
SMB100A
443Operating Manual 1407.0806.32 ─ 21
:SYSTem:COMMunicate:SERial:SBITs <SBits>
Sets the number of stop bits for the serial remote control interface.
Parameters:
<SBits> 1 | 2
*RST: 1
Example:
SYST:COMM:SER:SBIT 2
selects 2 stop bits.
Manual operation: See "Stop Bits" on page 110
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:SOCKet:RESource?
Response: "TCPIP::10.113.1.150::5025::SOCKET"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 110
:SYSTem:IDENtification <Identification>
Selects the mode the instrument identification is performed.
Parameters:
<Identification> AUTO | USER
AUTO
The "IDN String" and the "OPT String" are set automatically.
USER
Enables the selection of user definable "IDN String" and "OPT
String".
*RST: AUTO
Example:
SYST:IDEN USER
selects the user defined identification string.
Manual operation: See "Mode" on page 111
:SYSTem:LANGuage <Language>
Sets the remote control command set.
The instrument can also be remote controlled via the command set of several other
generators, for example HP generator. See the Application Note 1GP71 at the down-
load area of the product site on the Internet.
SYSTem Subsystem