Remote Control Commands
R&S
®
SMB100A
441Operating Manual 1407.0806.32 ─ 21
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command :SYSTem:COMMunicate:GPIB[:
SELF]:ADDRess.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:GPIB:RESource?
queries the VISA resource string.
Response: "GPIB::28::INSTR"'
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 110
:SYSTem:COMMunicate:NETWork:RESource?
Queries the VISA resource string, used for remote control of the instrument with
VXI-11 protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:NETWork:RESource?
Response: "TCPIP::192.1.2.3::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 110
:SYSTem:COMMunicate:HISLip:RESource?
Queries the VISA resource string, used for remote control of the instrument with
HiSLIP protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:HISLip:RESource?
Response: "TCPIP::192.1.2.3::hislip0::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 110
:SYSTem:COMMunicate:USB:RESource?
Queries the visa resource string for remote control via the USB interface.
Return values:
<Resource> string
SYSTem Subsystem