Remote Control Commands
R&S
®
SMB100A
371Operating Manual 1407.0806.32 ─ 21
Parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list in List mode is restarted as soon
as it is finished.
SINGle|BUS
The list is triggered by the GPIB commands [:SOURce<hw>]:
LIST:TRIGger:EXECute. The list is executed once.
EXTernal
The list is triggered externally via the INST TRIG connector. The
list is executed once.
*RST:
AUTO
Example:
LIST:TRIG:SOUR EXT
selects triggering by means of the external trigger.
Manual operation: See "Mode - List Mode" on page 189
7.13.8 SOURce:MODulation Subsystem
This subsystem contains the command for switching on/off all modulations.
[:SOURce<hw>]:MODulation[:ALL][:STATe] <State>
Activates/deactivates the modulations.
The command SOUR:MOD:ALL:STAT OFF switches all modulations off. A subsequent
command SOUR:MOD:ALL:STAT ON restores the status that was active before the
last switch-off. "MOD OFF" is displayed in the info line of the header next to the "Level"
field.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
MOD:STAT OFF
switches off all modulations.
Manual operation: See "MOD ON/OFF" on page 197
7.13.9 SOURce:PGEN Subsystem
This subsystem contains the commands for setting the pulse generator.
[:SOURce<hw>]:PGENerator:STATe............................................................................... 372
SOURce Subsystem