Remote Control Commands
R&S
®
SMB100A
286Operating Manual 1407.0806.32 ─ 21
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage: Event
Manual operation: See "Factory Preset" on page 120
7.4 CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is trig-
gered by the query commands. The response "0" indicates error-free adjustment, and
the response "1" means that an error occurred during adjustment.
:CALibration<hw>:ALL[:MEASure]?................................................................................. 286
:CALibration<hw>:FMOFfset[:MEASure]?.........................................................................286
:CALibration<hw>:FREQuency[:MEASure]?......................................................................287
:CALibration<hw>:LEVel[:MEASure]?...............................................................................287
:CALibration<hw>:LEVel:EXTern:DATA............................................................................287
:CALibration<hw>:ROSCillator[:DATA]............................................................................. 288
[:SOURce]:CALibration:STEReo:ANALog[:MEAS]?........................................................... 288
:CALibration<hw>:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not require external measurement equipment.
Query parameters:
<Force> string
Return values:
<Measure> select
Example:
CAL:ALL:MEAS?
// Response "0"
// Adjustment has been performed successfully
Usage: Query only
Manual operation: See "Adjust All" on page 93
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure> 0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage: Query only
CALibration Subsystem