Remote Control Commands
R&S
®
SMB100A
287Operating Manual 1407.0806.32 ─ 21
Manual operation: See "Adjust FM Offset" on page 205
:CALibration<hw>:FREQuency[:MEASure]?
Starts all adjustments which affect the frequency.
Return values:
<Measure> 0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage: Query only
Manual operation: See "Adjust Synthesis" on page 93
:CALibration<hw>:LEVel[:MEASure]?
[<Force>]
Starts all adjustments which affect the level.
The acquired correction values improve the settling time and the signal quality.
Query parameters:
<Force> string
*RST: force
Return values:
<Measure> 0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See "Adjust Level" on page 94
:CALibration<hw>:LEVel:EXTern:DATA <Data>
Queries what data has been used for the level calibration.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Parameters:
<Data> FACTory | CUSTomer
*RST: FACTory
CALibration Subsystem