Remote Control Basics
R&S
®
SMB100A
269Operating Manual 1407.0806.32 ─ 21
3. Poll the operation complete state periodically (by means of a timer) using the
sequence: *OPC; *ESR?
A return value (LSB) of 1 indicates that the overlapped command has finished.
6.5 Status Reporting System
The status reporting system stores all information on the current operating state of the
instrument, and on errors which have occurred. This information is stored in the status
registers and in the error queue.
You can query both with the commands of the STATus Subsystem.
6.5.1 Hierarchy of the Status Registers
The Figure 6-1 shows the hierarchical structure of information in the status registers
(ascending from left to right).
Status Reporting System